Official recall record
Temporomandibular Joint (TMJ) Packs
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-10-17
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Cardinal Health was recently informed of a medical device recall from Integra Lifesciences concerning the Codman® surgical patties referenced above. The recall is being issued due to higher-than-expected levels of endotoxin within the raw material used to produce Codman® surgical patties & surgical strips that may have resulted in out-of-specification levels of endotoxin in those finished goods. Recall start date: October 4, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.