OES Pro Urology Resection Sheaths (ref 76413)

Health Canada — Medical Devices · 2024-10-30 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-10-30
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Olympus has received six complaints about a damaged tip during use of a laser probe. Olympus has not received any reported injuries related to this matter. As a result of this complaint investigation, Olympus is removing the statement of compatibility with a greenlight laser for BPH therapy from the instructions for use (IFU) until Olympus can conduct additional testing. Recall Start Date: October 11, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.