Evis Exera III Video System - Gastrointestinal Videoscope (ref 76429)

Health Canada — Medical Devices · 2024-11-01 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-11-01
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Olympus previously issued revised cleaning instructions for the GIF-1TH190 on October 11, 2023, to issue new updated channel drying steps. Olympus conducted user evaluation, through human factors validation, and determined further clarification is required to improve the instructions for easier understanding of each drying step. As a result, Olympus is issuing a revised IFU addendum that informs customers that the channel drying time has been extended by 12 minutes to purge water from the channels. This is accomplished by drying each channel for 3 minutes, in the sequence described in the addendum. Ensure all reprocessing personnel are completely knowledgeable and thoroughly trained on the new water purge requirements. Olympus has received 16 reportable complaints related to this issue. Recall Start Date: October 18, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.