Official recall record
Vitros Chemistry Products System - Crp Slides (C-Reactive Protein)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-11-06
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
QuidelOrtho™ received complaints on the affected coating of VITROS® CRP slides, related to lower-than-expected results when testing patient samples known to have elevated concentrations of CRP (>90 mg/l or >9 mg/dl). When tested neat, results were reported as being within the upper end of the measuring range (70-90 mg/l or 7-9 mg/dl) however, upon dilution, results for these samples were reported as having concentrations greater than 2x the neat result. Patient samples with a CRP concentration above 90 mg/l (9 mg/dl) (typically above 150 mg/l (15 mg/dl)) may be reported as falsely reduced when samples are tested neat, with the dilute sample results predicting greater than 2x greater than the neat result. The observed bias will vary; however, the predicted neat CRP results will generally range between 70-90 mg/l (7-9 mg/dl). Recall start date: October 25, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.