ProductGuru Recalls

Official recall record

Medline Epidural Catheter

Health Canada — Medical Devices · 2024-11-07 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-11-07
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Medline Industries, LP has initiated a recall on convenience kits containing NEXUS™ Connectors A-EP-081 and A-EP-424 from Epimed International Inc. The connectors may not allow catheters to be inserted as intended if the connector is not in the fully open position. If the defect is not detected before use, there is a risk of medication leakage. Additionally, if the defect goes unnoticed, there is a risk of the catheter connector disconnecting during use. Recall start date: October 25, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.