Atellica CI System

Health Canada — Medical Devices · 2024-11-15 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-11-15
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Siemens Healthineers has confirmed through the internal investigation of customer complaints that the IMT diluent volume remaining (% remaining) does not decrease as expected on the Atellica® CI analyzer. The IMT module runs a standard a refresh sequence before processing samples if at least two minutes of inactivity have passed. IMT diluent is used by this sequence, however, the calculated diluent volume is not accurate, potentially leading to the IMT diluent being empty while still displaying that volume is remaining. In this case sodium (Na), potassium (K) and chloride (Cl) test results may be falsely elevated. quality control materials demonstrate the same behavior. Recall start date: November 6, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.