ProductGuru Recalls

Official recall record

Cardinal Health Level 3 Isolation Gowns

Health Canada — Medical Devices · 2024-11-20 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-11-20
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The listed product codes and all lots of Cardinal Health Level 3 Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (the Association for the Advancement of Medical Instrumentation) level 3 protection requirements. The raw material identified has demonstrated possible degradation due to exposure to lighting, temperature, and relative humidity beyond the manufacturer recommended limits. This product recall is specific to the codes listed. Other codes are not impacted. Recall Start Date: November 5, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.