TDHistoCyto Software (ref 76595)

Health Canada — Medical Devices · 2024-11-27 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-11-27
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Inappropriate behaviour was noted on a TDHC user site, specifically with regard to send/receipt by an external host system. In some rare cases, analysis reports were included in the hospital information system (HIS) with results from another patient. This recall applies only to systems with an INST0031 connection. Recall Start Date: August 28, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.