ProductGuru Recalls

Official recall record

Proteus XR/a General Purpose Radiographic System

Health Canada — Medical Devices · 2024-11-28 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-11-28
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

GE Healthcare has become aware that steel cables in the wall stand for certain Proteus XR/a systems serviced by GE Healthcare have not been replaced at the intervals detailed in the preventative maintenance action list (GEHC service manual 2273022-100). Recall Start Date: November 4, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters