Official recall record
Collar Button Drain Tube
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-11-29
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Olympus has identified via customer feedback that the carton packaging of two (2) specific lots; SM405666, JB362646, of Duravent Silicone Ventilation Tubes may contain incorrect product, specifically, 1.32mm Tiny T Tubes (model: 70240076) instead of the intended Duravent Silicone Ventilation Tubes. The inner sterile clamshell packaging correctly identifies the product within; however, it may not match the product identified on the carton label. No injuries or adverse events have been reported to Olympus as a result of the packaging error. Recall Start Date: November 19, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.