Access Intact PTH

Health Canada — Medical Devices · 2024-11-29 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2024-11-29
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Beckman Coulter has determined that 91 Access Intact PTH reagent packs (lot 339071) were filled with an incorrect volume of reagent in well 0. Access Intact PTH patient and QC samples with a reagent pack that has less volume than expected may: - Recover up to 30% lower than the expected results for routine PTH, with a pack imprecision of up to 10%. - Recover up to 40% lower than the expected results for intraoperative PTH, with a pack imprecision of up to 15%. - Produce a zero or near-zero result. Falsely decreased Intact PTH results could lead to a delay in diagnosis, additional diagnostic testing, and/or inappropriate treatment. Recall Start Date: November 25, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.