ProductGuru Recalls

Official recall record

EGM - Essential Gas Module

Health Canada — Medical Devices · 2024-12-05 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-12-05
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Philips has identified a potential safety issue for the 866173 IntelliVue G7m anesthesia gas modules, which could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, inop (inoperable) alarm. It was anticipated that the expected operating lifetime of the pump was at least 10000 hours. However, based on testing, the average operating lifetime was found to be 4500 hours. The pump may fail prematurely when there is friction caused by an abrasion. Recall restart date: February 14, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.