Official recall record
BD FACSLyric Flow Cytometer
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-12-10
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
BD has identified through complaints an increase in field failures / replacements of the power supply module on BD FACSLyric Flow Cytometers. The capacitor on the power supply can split which has the potential for the instruments to fail to power on and/or stay on. Recall Start Date: November 20, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.