BD FACSLyric Flow Cytometer

Health Canada — Medical Devices · 2024-12-10 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-12-10
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

BD has identified through complaints an increase in field failures / replacements of the power supply module on BD FACSLyric Flow Cytometers. The capacitor on the power supply can split which has the potential for the instruments to fail to power on and/or stay on. Recall Start Date: November 20, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.