Official recall record
ACCOLADE™ and VISIONIST™ CRT-P Pacemakers
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-12-16
- Last updated (per notice)
- 2025-08-22
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
Boston Scientific is initiating an advisory to provide information about a subset of pacemakers from the ACCOLADE™ family and VISIONIST™ CRT-Ps with an increased potential to initiate safety mode during telemetry or, in rare instances, other normal, higher-power operations due to latent, high battery impedance when the device reaches approximately four (4) years or less of remaining battery longevity. Recall start date: December 12, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.