ProductGuru Recalls

Official recall record

ACCOLADE™ and VISIONIST™ CRT-P Pacemakers

Health Canada — Medical Devices · 2024-12-16 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2024-12-16
Last updated (per notice)
2025-08-22
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Boston Scientific is initiating an advisory to provide information about a subset of pacemakers from the ACCOLADE™ family and VISIONIST™ CRT-Ps with an increased potential to initiate safety mode during telemetry or, in rare instances, other normal, higher-power operations due to latent, high battery impedance when the device reaches approximately four (4) years or less of remaining battery longevity. Recall start date: December 12, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters