Change Healthcare Cardiology Hemo

Health Canada — Medical Devices · 2024-12-18 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-12-18
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

An issue has been identified with the Change Healthcare Cardiology Hemo System. Under certain circumstances, when a user manually edits medication information, such as dosage, route, or units, the system may not consistently save these changes correctly. Consequently, a medication information discrepancy may occur between the procedure notes (PN) and the vitals and meds (WM) modules. As a result, the generated hemo procedure report may display inconsistencies between the PN and VM sections. Recall start date: December 2, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.