ProductGuru Recalls

Official recall record

Coloplast Catheters

Health Canada — Medical Devices · 2024-12-30 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-12-30
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

This recall is due to breach in sterility barrier (packaging pouch). The products involved with a sterilization issue was tested to evaluate impact to package integrity (blue-dye penetration test). Pouch integrity failures were observed due to holes in the pouch. This defect is not easily visible by the product's users. Recall start date: December 2, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters