Single Use Mechanical Lithotriptor V
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-12-31
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Olympus has identified an increase in complaints for BML-V442QR-30. The complaint data analysis found that distal tip tearing of the Mechanical Lithotriptor V had increased beginning with the production of lot 33k. Olympus has identified 296 complaints for the BML-V442QR-30 globally between June 1, 2021, through July 31, 2024. There were 169 reportable malfunctions, and there was one report of a serious injury in relation to this issue. Recall start date: December 18, 2024