HemosIL Heparin Calibrators

Health Canada — Medical Devices · 2025-01-02 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-01-02
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Instrumentation Laboratory has received complaints of heparin controls reporting low out of their assigned acceptance range with HemosIL Liquid Anti-XA, leading to delays in patient testing. Our internal investigation has identified that the value assignment of the HemosIL Heparin Calibrators has shown a bias against the who international standards for UFH and LMWH, potentially causing reduced recovery when measuring heparin on both controls and samples. Recall start date: December 18, 2024