HemosIL Heparin Calibrators
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-01-02
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Instrumentation Laboratory has received complaints of heparin controls reporting low out of their assigned acceptance range with HemosIL Liquid Anti-XA, leading to delays in patient testing. Our internal investigation has identified that the value assignment of the HemosIL Heparin Calibrators has shown a bias against the who international standards for UFH and LMWH, potentially causing reduced recovery when measuring heparin on both controls and samples. Recall start date: December 18, 2024