ProductGuru Recalls

Official recall record

Zeus® Infinity® Empowered

Health Canada — Medical Devices · 2025-01-07 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-01-07
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

In the course of service activities, it was found that CO2 readings were slightly increased. In some cases, the ISO accuracy was no longer achieved. The tolerance check is intended to ensure that the zeus ie will measure all gases within the specified accuracy until the next annual check. Draeger has updated the instructions for use of the Zeus® ie anesthesia device regarding the test tolerance of the CO2 measurement. Recall start date: December 23, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.