ProductGuru Recalls

Official recall record

Infinity® Centralstation Wide

Health Canada — Medical Devices · 2025-01-08 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-01-08
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The recall is being initiated because the new edifier speaker configured with the ICS in analog mode, may not work as expected after a power loss or power (usb cable) being disconnected. The new edifier speaker configured with ICS in analog mode will default back to usb mode, after power being restored or power cable (usb cable) reconnected which can lead to the loss of audio alarms. Recall start date: December 23, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.