Vitros® 5600, 3600 and XT Systems

Health Canada — Medical Devices · 2025-01-09 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-01-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

During internal testing, QuidelOrtho™ identified that microwell assay results have the potential to be reported without the reagent expired (re) result code, even when the signal reagent (sr) pack is past the expiry date. This can occur when an sr pack exceeds the on-board expiry and/or the shelf expiration date. This issue can affect any microwell assay regardless of whether or not the system is configured to use expired reagents. Recall start date: December 24, 2024