ProductGuru Recalls

Official recall record

ETHIBOND EXCEL® , ETHIBOND®, PERMAHAND®, PROLENE® sutures

Health Canada — Medical Devices · 2025-01-09 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-01-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

An open seal defect was detected during internal rework activities. Investigation determined that the preliminary cause of issue was a machine failure, leading to the zipper tray to be incorrectly positioned in the copolymer pre-formed cavity and subsequent tray misalignment with the front seal area in the sealing process. This ultimately resulted in possibility of open seal condition and/or narrow seal defects in the subsequent units. Recall start date: December 30, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.