ProductGuru Recalls

Official recall record

Heartware Ventricular Assist System - Heartware Monitor

Health Canada — Medical Devices · 2025-01-16 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-01-16
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

If steps for programming an HVAD controller are not followed correctly and a patient's driveline is connected to a controller that was not powered down after programming, the pump will not restart and could lead to patient harm. It is recommended to power down the programmed controller and disconnect the monitor data cable before performing a controller exchange. Recall start date: January 8, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.