AK 96 and AK 98 Dialysis Machines
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-01-17
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
Baxter Healthcare Corporation (Baxter) is issuing an urgent medical device correction for the AK 96 and AK 98 dialysis machines with the self-care configuration listed below due to the potential for hemolysis caused by the arterial dialyzer line kinking from the position of the remote panel arm. Recall start date: January 16, 2025