Official recall record
MEDRAD® Centargo Injection System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-01-28
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Centargo patient line sensor is a component of the Centargo injection system which detects a patient line. An investigation into reported air injection concerns found that, under certain special conditions, the patient line sensor may occasionally have difficulty detecting the presence of the patient line. This intermittent detection issue, combined with user reliance on automated system functions (such as auto-prime) and possible workflow inconsistencies, may slightly increase the likelihood of small air injections (1-10 ml). However, the overall risk remains low. Recall start date: December 13, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.