MEDRAD® Centargo Injection System

Health Canada — Medical Devices · 2025-01-28 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-01-28
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Centargo patient line sensor is a component of the Centargo injection system which detects a patient line. An investigation into reported air injection concerns found that, under certain special conditions, the patient line sensor may occasionally have difficulty detecting the presence of the patient line. This intermittent detection issue, combined with user reliance on automated system functions (such as auto-prime) and possible workflow inconsistencies, may slightly increase the likelihood of small air injections (1-10 ml). However, the overall risk remains low. Recall start date: December 13, 2024