MOSAIQ (ref 76878)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-01-29
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
MOSAIQ 2D image review application may provide incorrect image offsets when a 2D verification image is taken at a patient support angle (couch rotation) other than 0 degrees, and if the treatment machine is characterized at a scale that is not IEC61217. If the incorrect shift was not detected and used for treatment, mistreatment could occur. Recall start date: January 15, 2025