Official recall record
CIC-C1q - Circulating Immune Complexes (CIC) ELISA
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-01-30
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
In-house testing with CIC lot 3836 has revealed unexpectedly elevated values in internal control serum samples exceeding the expected target range. Notably, samples with target values below 5 μg eq/ml showed increased results of up to 50% above their target hence, values of samples around the recommended cut-off (5 μg eq/ml) obtained with this lot could incorrectly exceed the true value of the sample, and thus potentially lead to inaccurate conclusions. Recall start date: January 17, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.