Official recall record
VNS Therapy SenTiva Generators
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-02-05
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Complaints have been received where the SenTiva generator is reporting a loss of therapy or inability to make therapy changes with the VNS therapy programming system. A reed switch malfunction, resulting in the reed switch sticking in a closed position has been identified as the cause of these device events. Recall start date: January 10, 2025