Heater-Cooler Unit HCU 40

Health Canada — Medical Devices · 2025-02-06 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-02-06
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The ice bracket is attached to the inlet of the cardioplegia tank where ice is produced. To avoid interrupting the water circulation on the cardioplegia side, ice must not cover the opening at the bottom of the tank. Due to its very high thermal conductivity, the ice bracket prevents ice buildup in the affected area and the opening always remains patent. An internal investigation came to the conclusion that the failure rate of the ice bracket becomes unacceptably high after two years. For devices where the ice bracket was replaced more than 2 years ago: The ice bracket shall be replaced as soon as possible free of charge. For devices where the ice bracket was replaced less than 2 years ago: The ice bracket shall be replaced during the next preventive maintenance free of charge. In the future (not part of this field action): The ice bracket will be replaced biennially during preventive maintenance. Recall start date: January 24, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.