IntelePACS (Image Fusion Module)

Health Canada — Medical Devices · 2025-02-19 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-02-19
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

An incorrect SUV (standard uptake value) calculation occurs in the PET/CT fusion viewer module. This issue only occurs on days with a DST (daylight savings time) change. The issue occurs because the calculation relies on a time calculation which does not correctly consider the DST change. An incorrect SUV may lead to an incorrect diagnosis. Recall start date: February 4, 2025