Official recall record
Atellica CH Diluent
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-02-26
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Siemens Healthineers has confirmed, through the investigation of customer complaints, the presence of sodium hypochlorite (NaOCl) contamination in certain bottles of Atellica CH Diluent lot number 0000191602, bottle sequence numbers 00001-04400 (cartons 00001-02200) used on Atellica CH 930 and Atellica CI analyzers. Atellica IM assays are not affected. As a result of this issue, calibrator, quality control (QC), and patient results could be affected for any Atellica CH assay excluding sodium, potassium, and chloride. This issue is specific to Atellica CH Diluent lot number 0000191602. Unaffected alternate lots of this product are available. Recall start date: February 14, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.