Official recall record
Avalon US Transducer
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-03-12
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
The latest version of the wired Avalon ultrasound transducers (867246) may report inaccurate fetal heart rate (FHR) measurements when monitoring multiples (twins or triplets): in situations where the physiological signal (echo from the fetal heart) is absent or very weak (e.g. in early weeks of pregnancy), there is a tendency of the wired ultrasound transducers to interfere with each other and subsequently to produce an artificial FHR, mostly at approximately 180 bpm. The use cases affected by this issue are twin or triplet monitoring with wired ultrasound transducers including at least one transducer with software version l.01.04. The population at risk is women with multiple gestation (twins/triplets). Recall start date: March 10, 2025