ProductGuru Recalls

Official recall record

Comprehensive® Shoulder System Primary Mini Length Shouldler Stem

Health Canada — Medical Devices · 2025-03-13 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-03-13
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive® Shoulder Primary Stem (7mm Micro & 7 mm Mini). Five complaints reported that the stem implant did not fit in the inserter or that the inserter could not be detached from the stem. There were no injuries reported associated with the five complaints received. The internal investigation identified excess metal on the inner taper of the stem due to a chipped manufacturing tool. Recall start date: March 10, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.