ABL90 Flex and Flex Plus

Health Canada — Medical Devices · 2025-03-25 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-03-25
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

There is a potential risk with ABL90 Flex & ABL90 Flex Plus running RWIN10 1.0 and RWIN10 1.1. The issue is that the analyzer software may unexpectedly freeze during a sample measurement. The analyzer software includes a watchdog that detects a freeze and automatically restarts the software. If the analyzer freezes, a sample being processed will be lost. Recall Start Date: March 11, 2025