Stryker Instruments System 8 and 9 Saws
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-03-26
- Last updated (per notice)
- 2025-04-24
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
There is a potential for the compression spring (0038-419-000) inside the saw's drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade. Recall Start Date: March 14, 2025