Official recall record
WB Anatomic Healing Abutment XC
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-04-23
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Institut Straumann Ag received a complaint from a customer that the blister label for the WB XL Anatomic Healing Abutments says "RB/WB" instead of "WB", where "RB" and "WB" refer to the compatibility with the prosthetic connection (regular base and wide base). While the label on the outer box is correct, due to the incorrect blister label the use of the mislabeled WB abutments on RB implants cannot be excluded, in which case the abutment connection would exceed the implant shoulder. The affected devices were distributed from 2025-01-30 to 2025-03-13. Recall Start Date: April 7, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.