CADD-Solis™ Ambulatory Infusion Pump (ref 77345)

Health Canada — Medical Devices · 2025-04-28 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-04-28
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Issue 1: Damage to the circuit board within the housing of the rechargeable battery pack may cause melting of the plastic housing, specifically on the top and bottom surfaces. Issue 2: Damage to the wireless communication module circuit board may cause melting of the plastic housing of the wireless communication module battery. Issue 3: Damaged or dislodged battery separators, or foreign material in the battery compartment, may cause an electrical short condition between the battery contacts. Recall start date: April 8, 2025