Official recall record
CADD-Solis™ Ambulatory Infusion Pump (ref 77345)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-04-28
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Issue 1: Damage to the circuit board within the housing of the rechargeable battery pack may cause melting of the plastic housing, specifically on the top and bottom surfaces. Issue 2: Damage to the wireless communication module circuit board may cause melting of the plastic housing of the wireless communication module battery. Issue 3: Damaged or dislodged battery separators, or foreign material in the battery compartment, may cause an electrical short condition between the battery contacts. Recall start date: April 8, 2025