ProductGuru Recalls

Official recall record

Endopath Echelon Vascular Reloads

Health Canada — Medical Devices · 2025-05-02 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2025-05-02
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

An increase in firing issues complaints resulting from premature sled movement lockout (PMSLO) in the field. PMSLO is a safety mechanism that prevents the device from firing with a spent reload or no reload. It can accidentally be triggered if the cartridge is bumped and the sled moves out of position. Investigation has revealed that the force to bump the sled from the home position in the newest cartridge body tool is statistically lower than the previous tool, although there are no dimensional non-conformances. The IFU describes the steps to reduce risk of harm when using the device on critical vessels (i.e. proximal and distal control) and steps to safely remove the device if it locks on tissue. Recall start date: April 29, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.