Ortho Vision Analyzer

Health Canada — Medical Devices · 2025-05-09 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-05-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The loading station door unlocks for 20 seconds upon request to load samples or diluents. If the door is not opened within this period, it relocks without triggering a re-inventory since no user access occurred. If the door is opened immediately after the allotted 20-second countdown ends but within the door's auto-relock timeframe, the user will be allowed to load samples, however, re-inventory will not be triggered as the software does not detect that the loading station door was opened. There are no error codes generated for this issue. An inventory event will take place the next time you properly access the loading station. Recall Start Date: April 25, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.