Hugo RAS Tower, 120v

Health Canada — Medical Devices · 2025-05-09 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-05-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The loss of motion has been found to occur intraoperatively due to an instrument failure, causing the system to halt teleoperation with the robotic instrument. While the instrument is disabled and during the user's removal of the instrument it is possible for the user to trigger an error code by applying excessive torque which also disables the arm cart assembly (ACA). In both scenarios (disabled instrument, and disabled ACA) halting of teleoperation is an intended response of the system to prevent further use of the robotic instrument or arm cart assembly until the issue is resolved, or the portion of the system with issue is replaced. Recall Start Date: April 29, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.