ProductGuru Recalls

Official recall record

Midas Rex Legend, Burr-Tapered

Health Canada — Medical Devices · 2025-05-12 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-05-12
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

In October 2024, Medtronic became aware of a manufacturing issue involving Midas Rex Legend Fluted Spiral Tool that rendered the tools unusable. The cutting head of the F2/8TA23S tool is manufactured on a CNC grinder from a rod of material. The grinding process can leave excess material (burrs) as it grinds the cutting flutes into the stem. A deburring process downstream removes this material. It was determined that the process was not adequately performed for approximately six months in 2024. These burs can prevent the tool from being inserted into an attachment and therefore prevent the device from being used to perform a surgical procedure. Recall Start Date: April 30, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters