Official recall record
Affinity 4 Birthing Bed
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-05-22
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Baxter Corporation is issuing an urgent medical device correction for specific Affinity 4 Birthing Beds with hilo and foot actuators due to the potential for improper crimps on the signal wires of the actuators, which could result in a loss of ability to lower the bed or the foot section of the bed. Recall start date: May 7, 2025