Newport HT70 Ventilator Plus

Health Canada — Medical Devices · 2025-05-22 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2025-05-22
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

The C92 capacitor of HT70 ventilators may fail, disabling the output of the power management chip causing the ventilator to cease operation and stop ventilation. An audible alarm will be initiated, and the safety valve will open. In addition to the C92 issue, HT70 ventilators containing C159 capacitors reported failures of the shutdown buzzer alarm test, and as a result, the alarm "hardly sounded" or was of insufficient volume or duration per specification, placing a patient at risk of respiratory failure due to unexpected loss of ventilation. Recall start date: May 21, 2025