ProductGuru Recalls

Official recall record

Vantage Gastroscopes and Colonoscopes

Health Canada — Medical Devices · 2025-06-04 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-06-04
Last updated (per notice)
2025-07-16
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The devices were recently repaired at Vantage's service centre and returned to the customers. It came to their attention that the irrigation water pressure of these device(s) may exceed the specified range, potentially resulting in stronger-than-expected pressure during a procedure. Recall start date: May 5, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters