Alinity I Total PSA Reagent Kit

Health Canada — Medical Devices · 2025-06-05 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-06-05
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

An elevated number of complaints have been reported involving both out-of-range third-party quality control (qc) results and a positive bias in patient sample results when using specific product lots. Internal testing with patient samples, including retained lot analysis and trending of patient result medians, confirmed that some lots may exhibit a greater than 10% positive bias. Recall start date: May 22, 2025