Official recall record
Hamilton-C6 Intensive Care Ventilator
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-06-11
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Hamilton Medical AG received information from the supplier of the ESM-board which is installed in the Hamilton-C6, indicating that certain ESM-board may contain a mechanically damaged ceramic capacitor. Since only certain serial numbers (from 13636 till 13950 and from 13966 till 14753) of the Hamilton-C6 are affected with the defective ESM-board, this FSN is limited to these devices defined above. Damage to the capacitor could, over time, lead to a loss of its function due to oxidation or similar degradation processes. The worst-case scenario is a local short circuit of the aforementioned capacitor, with different outcomes depending on the device state. Recall start date: June 2, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.