BD FACS™ 7-Color Setup Beads

Health Canada — Medical Devices · 2025-06-11 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-06-11
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Based on customer complaints, BD has identified errors in the Spectral Overlap Factors (SOF) table provided for specific lots of BD FACS™ 7-color setup beads as listed in the table above. This issue may affect the accuracy of flow cytometry data due to incorrect flow cytometry setup compensation values for multicolor reagents for which proper fluorescence compensation is required prior to analyzing samples. Recall start date: June 3, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →