Pinnacle³ Radiation Therapy Planning System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-06-19
- Last updated (per notice)
- 2025-08-19
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
There is a software issue affecting Pinnacle³ Radiation Therapy Planning System performing advanced region of interest (ROI) expansion and contraction that results in a potential shift of the intended ROI. The software assumes the patient is always in the head first supine (HFS) position when transposing the advanced expansion to alternate positions. While performing advanced region of interest (ROI) expansion and contraction in the HFS position, the tool works as intended. However, when performing advanced ROI expansion and contraction on HFP, FFS, or FFP orientation images sets, a direction error occurs which results in potential overdose to nearby structure or potential underdose in the treatment plan. The following steps will lead to the software issue described above: Therapist / physician performs simulation based on HFP, FFS, and FFP images Physician performs advanced variable ROI expansion/contraction There is an error in direction of expansion/contraction Recall start date: June 11, 2025