INNOVANCE® VWF Ac assay

Health Canada — Medical Devices · 2025-07-04 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-07-04
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Siemens Healthineers has confirmed through the investigation of customer complaints that quality control and patient sample results for INNOVANCE® VWF Ac might be falsely elevated, independent of the system and reagent lot used. An increased number of complaints for the test INNOVANCE® VWF Ac were received, where customers complained of quality control (qc) recovery issues for control plasma p (cpp). Results were found above the permitted range according to the table of assigned values. There is a potential of elevated INNOVANCE® VWF Ac patient results where qc is within the permitted range. Date de début du rappel: le 20 juin 2025