Beckman Coulter Au Immunochemi Sys- Unsaturated Iron Binding Capacity (UIBC)

Health Canada — Medical Devices · 2025-07-09 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-07-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

UIBC OSR61205 is not meeting interference claims for haemoglobin interference serum application. As per product ifus the claim is less than 10% up to 2g/l haemoglobin. While the high pool is meeting the claim, the low pool is failing the hemolysate interference specifications with results of -43.63% or -8.95 μmol/l. Recall start date: July 1, 2025